Deep Learning Localizes and Identifies Polyps in Real Time With 96% Accuracy in Screening Colonoscopy

  • Gregor Urban
    Affiliations
    Department of Computer Science, University of California, Irvine, California

    Institute for Genomics and Bioinformatics, University of California, Irvine, California
    Search for articles by this author
  • Priyam Tripathi
    Affiliations
    Department of Medicine, University of California, Irvine, California
    Search for articles by this author
  • Talal Alkayali
    Affiliations
    Department of Medicine, University of California, Irvine, California

    H.H. Chao Comprehensive Digestive Disease Center, University of California, Irvine, California
    Search for articles by this author
  • Mohit Mittal
    Affiliations
    Department of Medicine, University of California, Irvine, California
    Search for articles by this author
  • Farid Jalali
    Affiliations
    Department of Medicine, University of California, Irvine, California

    H.H. Chao Comprehensive Digestive Disease Center, University of California, Irvine, California
    Search for articles by this author
  • William Karnes
    Affiliations
    Department of Medicine, University of California, Irvine, California

    H.H. Chao Comprehensive Digestive Disease Center, University of California, Irvine, California
    Search for articles by this author
  • Pierre Baldi
    Correspondence
    Reprint requests Address requests for reprints to: Pierre Baldi, PhD, 4038 Bren Hall, University of California, Irvine, California 92697-3435.
    Affiliations
    Department of Computer Science, University of California, Irvine, California

    Institute for Genomics and Bioinformatics, University of California, Irvine, California

    Center for Machine Learning and Intelligent Systems, University of California, Irvine, California
    Search for articles by this author

      Background & Aims

      The benefit of colonoscopy for colorectal cancer prevention depends on the adenoma detection rate (ADR). The ADR should reflect the adenoma prevalence rate, which is estimated to be higher than 50% in the screening-age population. However, the ADR by colonoscopists varies from 7% to 53%. It is estimated that every 1% increase in ADR lowers the risk of interval colorectal cancers by 3%–6%. New strategies are needed to increase the ADR during colonoscopy. We tested the ability of computer-assisted image analysis using convolutional neural networks (CNNs; a deep learning model for image analysis) to improve polyp detection, a surrogate of ADR.

      Methods

      We designed and trained deep CNNs to detect polyps using a diverse and representative set of 8,641 hand-labeled images from screening colonoscopies collected from more than 2000 patients. We tested the models on 20 colonoscopy videos with a total duration of 5 hours. Expert colonoscopists were asked to identify all polyps in 9 de-identified colonoscopy videos, which were selected from archived video studies, with or without benefit of the CNN overlay. Their findings were compared with those of the CNN using CNN-assisted expert review as the reference.

      Results

      When tested on manually labeled images, the CNN identified polyps with an area under the receiver operating characteristic curve of 0.991 and an accuracy of 96.4%. In the analysis of colonoscopy videos in which 28 polyps were removed, 4 expert reviewers identified 8 additional polyps without CNN assistance that had not been removed and identified an additional 17 polyps with CNN assistance (45 in total). All polyps removed and identified by expert review were detected by the CNN. The CNN had a false-positive rate of 7%.

      Conclusion

      In a set of 8,641 colonoscopy images containing 4,088 unique polyps, the CNN identified polyps with a cross-validation accuracy of 96.4% and an area under the receiver operating characteristic curve of 0.991. The CNN system detected and localized polyps well within real-time constraints using an ordinary desktop machine with a contemporary graphics processing unit. This system could increase the ADR and decrease interval colorectal cancers but requires validation in large multicenter trials.

      Graphical abstract

      Keywords

      Abbreviations used in this paper:

      ADR (adenoma detection rate), AUC (area under the curve), CNN (convolutional neural network), CRC (colorectal cancer), FPR (false-positive rate), NBI (narrow-band imaging), NPI (not pre-initialized), PI (pre-initialized), WLE (white light endoscopy)

       Background and Context

      The benefit of colonoscopy for colorectal cancer prevention depends on the adenoma detection rate (ADR). New strategies are needed to increase the ADR during colonoscopy.

       New Findings

      A system of convolutional neural networks (CNN) called Deep Learning was able to process colonoscopy images at high speed in real time, identifying polyps with a cross-validation accuracy of 96.4% and ROC-AUC value of 0.991.

       Limitations

      Possible effects of the CNN on inspection behavior by colonoscopists are not known. The anonymized videos excluded information about patient history. CNN performance may vary by indication (screening vs surveillance).

       Impact

      This technology may assist colonoscopists in finding precancerous polyps in real-time and with high accuracy.
      Colorectal cancer (CRC) is the second leading cause of cancer-related death in the United States.

      American Cancer Society. Cancer facts and figures 2016. Available at: https://www.cancer.org/research/cancer-facts-statistics/all-cancer-facts-figures/cancer-facts-figures-2016.html. Accessed September 3, 2018.

      CRC arises from precancerous polyps
      • Strum W.B.
      Colorectal adenomas.
      with a mean dwell time of at least 10 years.
      • Russell L.B.
      • Kuntz K.M.
      • Lansdorp-Vogelaar
      • et al.
      A systematic comparison of microsimulation models of colorectal cancer: the role of assumptions about adenoma progression.
      The National Polyp Study showed that 70%–90% of CRCs are preventable with regular colonoscopies and removal of polyps.
      • Winawer S.J.
      • Zauber A.G.
      • Ho M.N.
      • et al.
      Prevention of colorectal cancer by colonoscopic polypectomy.
      Seven percent of 9% of CRCs occur despite being up to date with colonoscopy.
      • Patel S.G.
      • Ahnen D.J.
      Prevention of interval colorectal cancers: what every clinician needs to know.
      It is estimated that 85% of these “interval cancers” are due to missed polyps or incompletely removed polyps during colonoscopy.
      • Pohl H.
      • Robertson D.J.
      Colorectal cancers detected after colonoscopy frequently result from missed lesions.
      The prevalence of precancerous polyps in the screening population older than 50 years is estimated to be at least 50%.
      • Leufkens A.M.
      • van Oijen M.G.H.
      • Vleggaar F.P.
      • et al.
      Factors influencing the miss rate of polyps in a back-to-back colonoscopy study.
      Adenomas are the most prevalent precancerous polyp. The adenoma detection rate (ADR; percentage of screening colonoscopies with ≥1 adenoma found) is a measure of a colonoscopist’s ability to find adenomas. Ideally, the ADR should reflect adenoma prevalence. Unfortunately, the ADR varies widely (7%–53%) among colonoscopists performing screening colonoscopies.
      • Corley D.A.
      • Jensen C.D.
      • Marks A.R.
      • et al.
      Adenoma detection rate and risk of colorectal cancer and death.
      In tandem colonoscopies, 22%–28% of polyps and 20%–24% of adenomas were missed
      • Leufkens A.M.
      • van Oijen M.G.H.
      • Vleggaar F.P.
      • et al.
      Factors influencing the miss rate of polyps in a back-to-back colonoscopy study.
      and CRC had a diagnostic miss rate of 5%.
      • Than M.
      • Witherspoon J.
      • Shami J.
      • et al.
      Diagnostic miss rate for colorectal cancer: an audit.
      The ADR is dependent on a colonoscopist’s level of training, time spent, and technique used during withdrawal, preparation quality, and other colonoscopist- and procedure-dependent factors.
      • Anderson J.C.
      • Butterly L.F.
      Colonoscopy: quality indicators.
      A large Kaiser Permanente study showed that for each 1% increase in the ADR, the interval cancer rate was decreased by 3%.
      • Corley D.A.
      • Jensen C.D.
      • Marks A.R.
      • et al.
      Adenoma detection rate and risk of colorectal cancer and death.
      A subsequent study with nearly 1 million person-years of follow-ups in Poland showed a 6% decrease in interval cancer rates for each 1% increase in the ADR.
      • Kaminski M.F.
      • Wieszczy P.
      • Rupinski M.
      • et al.
      increased rate of adenoma detection associates with reduced risk of colorectal cancer and death.
      This study also showed an 82% decrease in interval cancer rates among colonoscopists that improved their ADRs to the top quintile. Not surprisingly, the ADR currently is a key quality measure reportable in the United States to the Centers for Medicare and Medicaid and is tied to reimbursement under the Medicare Access and CHIP Reauthorization Act of 2015 and the Merit Based Incentive Payments System beginning in the 2017.

      GI Quality Measures for 2017 Released in MACRA Final Rule. Available at: http://partner.gastro.org/gi-quality-measures-for-2017-released-in-macra-final-rule. Accessed September 3, 2018.

      Several novel technologies have been developed to improve the ADR, including enhanced optics (resolution, zoom and wide angle, chromoendoscopy, digital autofluorescence, extra lenses for side and forward views) and attachments and modifications to aid view behind and between folds, including cap-assisted techniques and a balloon-assisted device.
      • Bond A.
      • Sarkar S.
      New technologies and techniques to improve adenoma detection in colonoscopy.
      Extra-wide angle colonoscopes and multi-camera systems initially showed promise to increase the ADR compared with standard forward-facing camera systems.
      • Bond A.
      • Sarkar S.
      New technologies and techniques to improve adenoma detection in colonoscopy.
      However, a recent meta-analysis and large randomized study showed no difference in the ADR compared with standard forward-viewing colonoscopy.
      • Hassan C.
      • Senore C.
      • Radaelli F.
      • et al.
      Full-spectrum (FUSE) versus standard forward-viewing colonoscopy in an organised colorectal cancer screening programme.
      A review of 5 studies on the effect of high-definition colonoscopes on the ADR showed conflicting evidence,
      • Bond A.
      • Sarkar S.
      New technologies and techniques to improve adenoma detection in colonoscopy.
      with 1 study concluding that the ADR is improved only for endoscopists with a low ADR (<20%).
      • Waldmann E.
      • Britto-Arias M.
      • Gessl I.
      • et al.
      Endoscopists with low adenoma detection rates benefit from high-definition endoscopy.
      Similarly, most studies on digital chromoendoscopy, specifically narrow-band imaging (NBI), have found that NBI does not improve the ADR compared with white light imaging.
      • Bond A.
      • Sarkar S.
      New technologies and techniques to improve adenoma detection in colonoscopy.
      Evidence suggests positive effects of autofluorescence, but it is associated with added expense and poor image resolution.
      • Bond A.
      • Sarkar S.
      New technologies and techniques to improve adenoma detection in colonoscopy.
      Computer-assisted image analysis has the potential to further aid adenoma detection but has remained underdeveloped. A notable benefit of such a system is that no alteration of the colonoscope or procedure is necessary.
      Deep learning has been successfully applied to many areas of science and technology,
      • Schmidhuber J.
      Deep learning in neural networks: an overview.
      such as computer vision,
      • Baldi P.
      • Chauvin Y.
      Neural networks for fingerprint recognition.
      • Simonyan K.
      • Zisserman A.
      Very deep convolutional networks for large-scale image recognition.
      • He K.
      • Zhang X.
      • Ren S.
      • et al.
      Deep residual learning for image recognition.
      • Redmon J.
      • Divvala S.
      • Girshick R.
      • et al.
      You only look once: unified, real-time object detection.
      • Szegedy C.
      • Liu W.
      • Jia Y.
      • et al.
      Going deeper with convolutions.
      speech recognition,
      • Graves A.
      • Mohamed A.R.
      • Hinton G.
      Speech recognition with deep recurrent neural networks.
      natural language processing,
      • Wu Y.
      • Schuster M.
      • Chen Z.
      • et al.
      Google’s neural machine translation system: bridging the gap between human and machine translation.
      games,
      • Silver D.
      • Huang A.
      • Maddison C.
      • et al.
      Mastering the game of Go with deep neural networks and tree search.
      • Wu L.
      • Baldi P.
      Learning to play GO using recursive neural networks.
      particle physics,
      • Baldi P.
      • Sadowski P.
      • Whiteson D.
      Searching for exotic particles in high-energy physics with deep learning.
      • Shimmin C.
      • Sadowski P.
      • Baldi P.
      • et al.
      Decorrelated jet substructure tagging using adversarial neural network.
      organic chemistry,
      • Fooshee D.
      • Mood A.
      • Gutman E.
      • et al.
      deep learning for chemical reaction prediction.
      and biology,
      • Di Lena P.
      • Nagata K.
      • Baldi P.
      Deep architectures for protein contact map prediction.
      • Wang J.
      • Fang Z.
      • Lang N.
      • et al.
      A multi-resolution approach for spinal metastasis detection using deep siamese neural networks.
      • Wang J.
      • Ding H.
      • Azamian F.
      • et al.
      Detecting cardiovascular disease from mammograms with deep learning.
      • Chang P.
      • Su L.
      • Baldi P.
      • et al.
      Deep learning convolutional neural networks accurately classify genetic mutations in gliomas.
      • Esteva A.
      • Kuprel B.
      • Novoa R.
      • et al.
      Dermatologist-level classification of skin cancer with deep neural networks.
      • Baldi P.
      Deep learning in biomedical data science.
      to name just a few areas and examples. A convolutional neural network (CNN) is a type of deep learning model that is highly effective at performing image analysis.
      Ideally, a polyp-detection assistance module should have a sensitivity of 1 (or close to it) to avoid false-negative results, but this comes at the cost of an increased false-positive rate (FPR) when the area under the curve (AUC; Performance Evaluation and Metrics section in the Supplement) is not close to 1. A large FPR, even with perfect sensitivity, diminishes the benefits of an assistance system when user desensitization comes into play. A polyp-detection module also must process images at a minimum of 30 frames per second to be applicable during colonoscopy. Therefore, surmounting the constraints of accuracy and processing speed were our primary goals.

      Methods

       Convolutional Neural Networks

      We trained different CNN architectures in this study, including models with weights initialized by training on the ImageNet data corpus,
      • Russakovsky O.
      • Deng J.
      • Su H.
      • et al.
      ImageNet large scale visual recognition challenge.
      before refining the weights in our dataset. All trained CNNs consisted of the same fundamental building blocks, including (1) convolutional layers, (2) fully connected layers, (3) maximum or average pooling, (4) nonlinear activation functions, and (5), optionally, batch normalization operations
      • Ioffe S.
      • Szegedy C.
      Batch normalization: accelerating deep network training by reducing internal covariate shift.
      and skip connections.
      • He K.
      • Zhang X.
      • Ren S.
      • et al.
      Deep residual learning for image recognition.
      • Szegedy C.
      • Vanhoucke V.
      • Ioffe S.
      • et al.
      Rethinking the inception architecture for computer vision.
      We followed each convolutional layer by the rectified linear (ReLu) activation function. The last hidden layer of the models was densely connected to the output units. For regression problems (localization), we optimized the L2 loss with linear output units. For classification (detection), we used softmax output units and optimized the Kullback–Leibler divergence. An overview of the different neural network layer types is presented in the Neural Network Architectures section in the Supplementary Material while Supplementary Table 1 presents the used CNN architectures.
      All experiments were implemented using the Keras
      • Chollet F.
      Keras.
      and Tensorflow

      Abadi M, Agarwal A, Barham P, et al. TensorFlow: Large-scale machine learning on heterogeneous systems. Available at: https://www.tensorflow.org/. Published 2015.

      software libraries.

       Model Regularization

      We used established techniques to decrease overfitting when training neural networks. We applied dropout
      • Hinton G.E.
      • Srivastava N.
      • Krizhevsky A.
      • et al.
      Improving neural networks by preventing co-adaptation of feature detectors.
      • Baldi P.
      • Sadowski P.
      The dropout learning algorithm.
      with a rate of 0.5 to the input of the first and second fully connected layers in all models. Prior studies have found that data augmentation improves deep learning performance,
      • Chatfield K.
      • Simonyan K.
      • Vedaldi A.
      • et al.
      Return of the devil in the details: delving deep into convolutional nets.
      a process of synthetically generating additional training examples by using random image transformations, including rotations and mirroring of the input images during the training process. Doing so forces the model to learn to become invariant to these transformations. We used random horizontal and vertical mirroring, rotations in the full range of 0°–90°, and shearing. Another technique we used to decrease overfitting was “early stopping,” in which a small subset of the training set is reserved exclusively for monitoring the CNN’s accuracy during training and the weights of the network at the point of best performance are saved, as opposed to the weights obtained at the end of training.

      Experiments

       Datasets and Preprocessing

      Five different datasets were used for training and/or evaluating the deep learning models presented in this work: (1) the general-purpose computer-vision ImageNet challenge
      • Russakovsky O.
      • Deng J.
      • Su H.
      • et al.
      ImageNet large scale visual recognition challenge.
      dataset was implicitly used to pre-train the model weights; (2) 8,641 hand-selected colonoscopy images from more than 2,000 patients were used to avoid a possible intrapatient polyp similarity bias; (3) a separately collected dataset of 1,330 colonoscopy images from different patients; (4) 9 colonoscopy videos; (5) a combined dataset consisting of the 8,641 images and 44,947 image frames extracted from the 9 videos; and (6) a separate dataset of 11 deliberately more “challenging” colonoscopy videos. All colonoscopy images and videos were recorded using state-of-the-art Olympus PCF-H190 colonoscopes (Olympus Corp, Center Valley, PA) with a resolution of 1,280 × 1,024 pixels and NBI capability. The combined duration of all 20 videos was approximately 5 hours (∼500,000 frames), with significant variation in the number of polyps per video (See Supplementary Table 2, Supplementary Table 3, Supplementary Table 4 and especially Supplementary Table 5 in this context).
      • 1.
        The ImageNet challenge dataset contains 1.2 million natural images of objects, such as boats, cars, and dogs, but no medical images. We reasoned that many fundamental features learnable on this dataset could be transferable to the task of detecting polyps and thus used it to pre-initialize the weights of some of our deep neural networks to test this hypothesis.
      • 2.
        The set of 8641 colonoscopy images contained 4,088 images of unique polyps of all sizes and morphologies and 4553 images without polyps (ie, the dataset was almost perfectly balanced; Figure 1). The dataset included white light and NBI images (Figure 1) and covered all portions of the colorectum, including retro-views in the rectum and cecum, appendiceal orifice, and ileocecal valve. The total number of NBI images was 840, with the remaining 7801 obtained in white light endoscopy (WLE) conditions. We deliberately and randomly included features such as forceps, snares, cuff devices, debris, melanosis coli, and diverticula in polyp- and non–polyp-containing images in a balanced fashion to prevent the machine learning system from associating the appearance of tools with the presence of polyps. The images were stored at a resolution of 640 × 480 pixels. Locations and dimensions of bounding boxes were recorded for images containing polyps by a team of colonoscopists (fellows and faculty at the University of California–Irvine with an ADR > 45% and >100 procedures).
        Figure thumbnail gr1
        Figure 1Examples of dataset. (Top row) Images containing a polyp with a superimposed bounding box. (Bottom row) Non-polyp images. Three pictures on the left were taken using NBI and 3 pictures on the right include tools (eg, biopsy forceps, cuff devices, etc) that are commonly used in screening colonoscopy procedures.
      • 3.
        A separate set of 1330 colonoscopy images (672 unique polyp and 658 non-polyp images) was collected from different patients.
      • 4.
        Colonoscopy videos were recorded and evaluated at their original resolution of 1280 × 1024 pixels. The first set of 9 videos was selected randomly from archived video studies.
      • 5.
        The larger dataset was obtained by augmenting the original set of 8641 images with 44,947 image frames selected from the 9 videos that were labeled as part of the first validation study. Because consecutive frames are highly correlated, we selected every eighth image frame that contained no polyp and every fourth image frame containing a polyp (resulting in 13,292 polyp frames and 31,655 non-polyp frames).
      • 6.
        Colonoscopy procedures of the second set of 11 videos were performed by a highly skilled colonoscopist (ADR ≥ 50%) and contained segments for which the scope was deliberately withdrawn without closing in on already identified polyps to mimic a missed-polyp scenario. This set of videos was used only for validation purposes in our experiments and never for training.
      All images from the different datasets were preprocessed identically before being passed to the machine learning models. As a first step, the individual frames were rescaled to a fixed size of 224 × 224 (unless noted otherwise). Then, the values of the pixels in each frame were normalized to be unit normally distributed by subtracting the mean pixel value from all pixels in the frame and dividing the resulting values by the standard deviation measured across all pixels. This preprocessing and normalization approach allowed us to apply the same neural network to data from different sources, with different resolutions, without requiring further adjustments.

       Neural Network Architectures and Training

      A detailed description of the deep neural network architectures, training methods, and algorithms is presented in the Architectures, Training, and Evaluation section in the Supplement. A short summary is given below.
      We trained and evaluated polyp-detection and -localization models separately for clarity and to exclude confounding factors. They were architecturally identical except for the final layer, which performed binary classification (detection) or regression (localization). We implemented localization by predicting the size and location of a bounding box that tightly enclosed any identified polyps. This allowed us to build CNNs that could operate in real time as opposed to prior attempts, based on individual pixel classification, that struggled to operate in real time.
      • Bernal J.
      • Sánchez J.
      • Vilarino F.
      Towards automatic polyp detection with a polyp appearance model.
      We tested 3 variations for training the polyp localization model: (1) optimizing the size and location with the mean-squared error loss (L2); (2) optimizing the Dice loss, which directly maximizes the overlap between the predicted bounding box and the ground truth; and (3) a variation of the “you only look once”
      • Redmon J.
      • Divvala S.
      • Girshick R.
      • et al.
      You only look once: unified, real-time object detection.
      algorithm in which the CNN produces and aggregates typically more than 49 individual weighted predictions of polyp size and location in a single forward pass; we refer to it as the “internal ensemble.” All detection and localization variants had almost identical run-time complexity (<1% difference).
      We experimented with several different CNN architectures (Supplementary Table 1) that fell into 1 of 2 categories: those initialized with random weights (denoted NPI for “not pre-initialized”) and those pre-initialized (PI) with weights obtained by training the corresponding model on the ImageNet Challenge data of natural images. Within the class of PI architectures, we used architectures VGG16,
      • Simonyan K.
      • Zisserman A.
      Very deep convolutional networks for large-scale image recognition.
      VGG19,
      • Simonyan K.
      • Zisserman A.
      Very deep convolutional networks for large-scale image recognition.
      and ResNet50.
      • He K.
      • Zhang X.
      • Ren S.
      • et al.
      Deep residual learning for image recognition.
      All CNN architectures (NPI and PI) were trained using the colonoscopy data. All experiments were performed using modern TITAN X (Pascal) graphics processing units (NVIDIA, Santa Clara, CA) with 12 GB of RAM and a processing power of 11 TFLOPS.

       Training and Testing

      Throughout the experiments, we used multiple splits of the same dataset (cross-validation) or trained on 1 dataset and tested the model on a completely different dataset. For early stopping and hyper-parameter optimization, we always set aside a small subset of the training set for monitoring the model’s performance. The main experiments were:
      • Cross-validation on the 8641 images
      • Training on the 8641 images and testing on the 9 videos, 11 videos, and independent dataset
      • Training on the 8641 images and 9 videos and testing on the 11 videos and independent dataset
      In most cases we use models pre-trained on ImageNet.

       Colonoscopy Video Study With Expert Colonoscopists

      Three expert colonoscopists (ADR ≥ 50%) were tasked to identify all polyps in 9 de-identified colonoscopy videos selected from the archived video studies without benefit of the CNN overlay. Experts recorded the first and last frames in which they believed they encountered a polyp in the videos. Their “polyp encounters” were combined by consensus. We filtered the CNN predictions for polyps by requiring at least 8 contiguous video frames with greater than 40% probability for polyp presence as predicted by the PI-CNN2 model alone. The decision to filter CNN predictions in blocks of 8 frames yielded a balanced sensitivity and specificity (see Supplementary Figure 1 for an analysis of different block sizes). This optimal setting might shift for a different training dataset or CNN model.
      We generated CNN-overlaid videos by superimposing a small green box on each frame where a polyp was detected with greater than 95% predicted probability at the location and with dimensions that were predicted by our polyp-localization CNN.
      A senior expert (ADR ≥ 50%, >20,000 colonoscopies) was tasked to review the CNN-overlaid videos and assign the uniqueness of each polyp and the confidence level of true polyp presence (high vs low; Figure 2). Contingency analysis compared the number of agreements and disagreements on individual video frames between post-processed CNN predictions alone and CNN-assisted expert review. We repeated this study using a second set of 11 more challenging videos (Datasets and Preprocessing section).
      Figure thumbnail gr2
      Figure 2Representative frame shots of CNN-overlaid colonoscopy videos. Presence of a green box indicates that a polyp is detected with greater than 95% confidence by our CNN polyp localization model; the location and size of the box are predictions of the CNN model. Expert confidence that the box contained a true polyp is shown in the upper left of the images (video collages of CNN localization predictions available at: http://www.igb.uci.edu/colonoscopy/AI_for_GI.html).

      Results

       Polyp Detection

      Polyp detection results are presented in Table 1 (see Supplementary Figure 2 for the corresponding ROC plots). The first 2 rows (NPI-CNN1 and -2) correspond to models that were trained starting from random weights and these obtained accuracies that were comparable to previously published state-of-the-art polyp classification models.
      • Ribeiro E.
      • Uhl A.
      • Häfner M.
      Colonic polyp classification with convolutional neural networks.
      Table 1Summary of Polyp Classification Results for Architectures (Supplementary Table 1) Obtained by 7-Fold Cross-Validation on 8,641 Colonoscopy Images
      ModelInitial weightsAccuracyAUCSensitivity at 5% FPRSensitivity at 1% FPR
      NPI-CNN191.9 ± 0.2%0.970 ± 0.00288.1%65.4%
      NPI-CNN291.0 ± 0.4%0.966 ± 0.00286.2%60.6%
      PI-CNN1VGG1695.9 ± 0.3%0.990 ± 0.00196.9%87.8%
      PI-CNN2VGG1996.4 ± 0.3%0.991 ± 0.00196.9%88.1%
      PI-CNN3ResNet5096.1 ± 0.1%0.990 ± 0.00196.8%88.0%
      NOTE. The sensitivity (true-positive rate) is given at false-positive rates of 5% and 1% (ie, a specificity of 95% and 99%, respectively), because sensitivity and specificity are interdependent values.
      FPR, false-positive rate.
      Networks pre-initialized from prior training on the large ImageNet dataset of natural images surpassed those starting from random weights by a significant margin (PI-CNN1–3), despite meticulous hyper-parameter optimization of all models. We also observed that the scores of the PI ImageNet models were surprisingly similar for the different architectures.
      At a sensitivity level (true-positive rate) of 90%, the best model (PI-CNN2) had an FPR of 0.5%; at a sensitivity of 95.2%, the FPR was 1.6%; and at 97.1% sensitivity, the FPR increased to 6.5%. Thus, there was a reasonably large range of high sensitivities at which the number of expected false-positive alerts remained very low.
      Nonpolypoid (flat and depressed) polyps are challenging to detect compared with polypoid polyps and were often neglected until their potential to cause CRC worldwide (beyond Japan only) was discovered.
      • Soetikno R.
      • Friedland S.
      • Kaltenbach T.
      • et al.
      Nonpolypoid (flat and depressed) colorectal neoplasms.
      To investigate whether the CNN could correctly classify all polyps irrespective of their morphology, we reviewed a random subset of 1578 true-positive polyp predictions (of 3860) and all 228 false-negative polyp predictions from the validation set of 8641 images. We categorized them using the Paris classification and their estimated size. The results are presented in Table 2 and show that 381 nonpolypoid lesions (IIa, IIb, and IIc) and 678 polypoid polyps (Ip and Is) were present in this subset. The CNN missed 12% of polypoid polyps (84 of 678) and 11% of nonpolypoid lesions (41 of 381) in this biased subset. Based on this similarity, we conclude that the CNN can detect either polyp type equally well. Furthermore, if we correct for the sampling bias by including all the remaining true-positive values, then the percentage of missed polyps decreases to approximately 5%.
      Table 2Categorization of a Random Subset of 1,578 True-Positive and All 228 False-Negative Polyp CNN Predictions on the Test Set of 4,088 Unique Polyps, Categorized by Size and Paris Classification
      True positivesFalse negatives
      ≤1 cm>1 cm≤1 cm>1 cm
      Diameter (≤3 mm)644103
      Ip372586
      Is48745682
      IIa24637364
      IIb341510
      IIc4400
      NOTE. Results were obtained by 7-fold cross-validation on the 8,641 colonoscopy images. All polyps larger than 3 mm are categorized by the Paris classification scheme. The CNN performs equally well at detecting nonpolypoid lesions (IIa–c) and polypoid polyps (Ip and Is).
      All previously discussed results were obtained with models operating on inputs scaled to 224 × 224 pixels, which is the native resolution for which VGG16, VGG19, and ResNet50 were designed. We also optimized and trained the models at a resolution of 480 × 480 pixels to test whether they would be limited by the lower resolution. In a 7-fold cross-validation experiment, the networks pre-initialized with weights from the VGG16, VGG19, and ResNet50 models yielded test accuracies of 96.4%, 96.1%, and 96.4% and AUCs of 0.990, 0.991, and 0.992, respectively. These results were almost identical (up to noise) to those obtained with a lower resolution of 224 × 224 pixels but more than double the computational cost (processing time).
      The VGG-19–based CNN trained on the 8,641 images was tested on the independent dataset of 1,330 images, where it achieved a test accuracy of 96.4% and an AUC of 0.974. This accuracy was identical to the accuracy obtained by cross-validation analysis on the 8641 colonoscopy images, further confirming that intrapatient polyp similarity does not present a notable bias.

       Polyp Localization

      We trained models on the polyp-localization task on the data subset containing only a single polyp per frame, which represents the vast majority of samples with polyps.
      The test set results, as presented in Table 3, showed that the PI ImageNet CNNs (PI-CNN1 and -2) performed significantly better at localizing polyps than the randomly initialized neural network, which is consistent with our findings for polyp-presence detection (Polyp Detection section).
      Table 3Summary of Polyp Localization Results for a Subset of Architectures (Supplementary Table 1) Obtained by 7-Fold Cross-Validation on 8,641 Colonoscopy Images
      ModelL2 regression (Dice)Dice score optimization (Dice)“Internal ensemble” regression (Dice)
      NPI-CNN10.63 ± 0.010.681 ± 0.0020.731 ± 0.006
      PI-CNN10.77 ± 0.010.76 ± 0.010.823 ± 0.003
      PI-CNN20.79 ± 0.010.784 ± 0.0040.827 ± 0.003
      NOTE. Standard deviation of mean estimate from the cross-validation is shown.
      Neither of the 2 loss functions (L2 vs Dice) showed a consistent advantage over the other.
      Further, we found that the “internal ensemble” was noticeably better than the other approaches by improving the Dice coefficient from 0.79 to 0.83 for the best model (PI-CNN2 and VGG19).
      To provide a sense of scale, in previously published literature,
      • Bernal J.
      • Sánchez J.
      • Vilarino F.
      Towards automatic polyp detection with a polyp appearance model.
      a Dice score of 0.55 was obtained on a polyp-segmentation task using a different dataset.

       Colonoscopy Video Study With Expert Colonoscopists

      Of 9 colonoscopy videos that we considered in this study, 36 polyps were identified by 3 experts reviewing unaltered videos and 45 were identified by reviewing CNN-overlaid videos. Only 28 of 45 polyps were removed by the original colonoscopists (Table 4). No unique polyps were missed by the CNN. Of the 9 additional polyps found with CNN assistance, confidence was high for 3 and low for 6 (cf Figure 2). The sensitivity and specificity of CNN predictions compared with expert review of CNN-overlaid videos (based on single frames) were 0.93 and 0.93, respectively (P < .00001 by χ2 test; Supplementary Table 3). Rare false-negative results were enriched with views of distant and field-edge polyps. False-positive results were enriched with near-field collapsed mucosa, debris, suction marks, NBI, and polypectomy sites. Tables 4 and 5 present polyp-level summaries of the results of the video study.
      Table 4Unique Polyps Found and Removed During Colonoscopy, Found by Expert Review, and Found by CNN-Assisted Expert Review of 9 Videos
      Polyp size (mm)Original colonoscopist (polyps removed)Expert reviewCNN-assisted review
      1–3121924
      4–6121316
      7–9001
      >10444
      Total polyps found283645
      NOTE. The VGG-19–based CNN was trained on the 8,641 colonoscopy images and applied to the 9 videos without further adaptation.
      Table 5Analysis of VGG-19–Based CNN Tested on All 20 Videos After Training on 8,641 Colonoscopy Images
      9 Videos11 “Challenging” videos
      “Challenging videos” were produced by an expert colonoscopists who performed “flybys” that included but avoided inspection of known polyps to mimic missed-polyp scenarios.
      Total polyps found4568
      Total polyps missed05
      Total false positives8146
      Average delay to detection (s)
      Delay to detection of a polyp is the time span from a polyp entering the field of view of the camera for the first time to the CNN producing its first positive polyp prediction. All false-positive findings with duration of at least 1 second are counted. A frame-by-frame analysis of the video study is presented in Supplementary Table 3.
      0.2 ± 0.11.3 ± 0.3
      NOTE. CNN-assisted expert review annotations were used as reference.
      a “Challenging videos” were produced by an expert colonoscopists who performed “flybys” that included but avoided inspection of known polyps to mimic missed-polyp scenarios.
      b Delay to detection of a polyp is the time span from a polyp entering the field of view of the camera for the first time to the CNN producing its first positive polyp prediction. All false-positive findings with duration of at least 1 second are counted. A frame-by-frame analysis of the video study is presented in Supplementary Table 3.
      The second set of 11 videos contained 73 unique polyps found by expert review with and without CNN assistance. The CNN trained on the 8,641 images identified 68 of 73 polyps at a frame-by-frame FPR of 7% compared with expert labels of the videos. Fine tuning the CNN on the labeled frames from the first video study enabled the CNN to identify 67 of 73 polyps at an FPR of 5% (or, on a more sensitive setting, 72 of 73 polyps at an FPR of 12%) in this second set of videos. The additional training samples from the videos of the first study noticeably helped decrease the number of false-positive detections, most likely because of the great abundance and variety of random artifacts such as water, air bubbles, fecal matter, and low-quality and blurry frames from quick movement. This suggests that using additional training data could lead to further improvements.
      The Supplement contains a detailed breakdown of the polyps found in the 20 videos by the experts, including their location, size, morphology, and other details.

       Additional Experiments

      We performed further experiments to test (1) whether a system trained on WLE and NBI could perform well on colonoscopies without NBI capability and (2) whether a system trained on WLE data could perform even better on WLE-only colonoscopies than a system trained using NBI and WLE. We retrained the same VGG-19–based CNN on NBI-only, WLE-only, and subsets of the 8,641 colonoscopy images in 7-fold cross-validation analyses. Training and testing the CNN on WLE-only data resulted in a cross-validation test accuracy of 96.1% and an AUC of 0.991, whereas training and testing on NBI-only data yielded an accuracy of 92.9% and an AUC of 0.970. In these cases, this was worse than what the same CNN achieved when trained on WLE plus NBI data: an accuracy of 96.4% with an AUC of 0.992 on WLE-only data and an accuracy of 94.8% and an AUC of 0.988 on NBI-only data. The test accuracy on NBI images was consistently worse than on WLE data, but this can be explained by the significantly smaller amount of NBI training data (840 NBI images vs 7,801 WLE images). In summary, we found that it was beneficial to train the CNN on NBI plus WLE images, because this increased the total amount of training data, resulting in a synergistic knowledge “transfer” between the 2 modalities.

      Discussion

      An early application of computer-assisted polyp detection used traditional non-learning–based computer-vision engineering methods and achieved an AUC of 0.98 for detecting a limited class of polyp types
      • Bernal J.
      • Sánchez J.
      • Vilarino F.
      Towards automatic polyp detection with a polyp appearance model.
      but could not work in real time, requiring 19 seconds to process a single frame. Of 8 submissions to the MICCAI 2015 Endoscopic Vision Challenge for polyp detection, none could operate in real time, and the most accurate (CUMED) had a detection accuracy of 89% tested across 18,092 video frames.
      • Bernal J.
      • Sánchez J.
      • Vilarino F.
      Towards automatic polyp detection with a polyp appearance model.
      Other CNNs applied to the task of polyp detection have been limited by small databases of polyp images and videos. An implementation operating on multiple small sub-patches of images reached a classification accuracy of 91%,
      • Ribeiro E.
      • Uhl A.
      • Häfner M.
      Colonic polyp classification with convolutional neural networks.
      whereas another approach using 3 convolutional layers operating on heavily subsampled images of 32 × 32 pixels obtained an AUC of 0.86, a sensitivity of 86%, and a specificity of 85%.
      • Park S.Y.
      • Sargent D.
      Colonoscopic polyp detection using convolutional neural networks.
      In this study we trained state-of-the-art CNNs, pre-initialized on millions of labeled natural images (ImageNet), on a dataset of more than 8,000 labeled colonoscopy images from more than 2,000 patients. For the first time, these models could identify and locate polyps in real time and simultaneously achieve high accuracy and AUC. The highest performing model could process 98 images per second (10 ms per frame) for polyp detection and localization when using a modern consumer-grade graphics processing unit. This is approximately 4 times faster than required to implement real-time assistance software, because commonly used video encodings are standardized to 25 or 30 frames per second (PAL and NTSC standards). In comparison, the fastest polyp localization model presented in prior work,
      • Bernal J.
      • Tajkbaksh N.
      • Sánchez F.
      • et al.
      Comparative validation of polyp detection methods in video colonoscopy: results from the MICCAI 2015 Endoscopic Vision Challenge.
      among 8, could process only 7 frames per second and the slowest one could process merely 0.1 frame per second. A main contribution to the speed of our system comes from the choice of locating polyps through bounding boxes rather than unnecessarily precise polyp boundaries.
      For resolution, although all human assessments of videos were performed at high resolution, we found that a 224- × 224-pixel resolution was sufficient for the CNNs and virtually indistinguishable from a 480- × 480-pixel resolution. We believe that future advances in computer hardware and machine learning will allow the use of even larger input fields while preserving real-time capabilities and potentially improving detection accuracy.
      Our feasibility study of 9 colonoscopy videos, reviewed by expert colonoscopists with and without the aid of a CNN overlay, showed that the model identified all polyps and aided discovery of additional polyps with relatively a low burden of false-positive results. In a second study, with 11 purposefully difficult colonoscopy videos, recorded by a senior colonoscopist, featuring “flyby” scenarios without closing in on previously found polyps during withdrawal, the CNN alone identified 67 of 73 unique polyps, with an average of fewer than 5 false-positive results per video. Missed polyps were located in “flyby” segments of the video, suggesting that CNN assistance cannot compensate for hurried withdrawal and poor inspection technique. Polyp morphology did not play a role in CNN performance, that is, nonpolypoid lesions were not missed by the CNN more often than polypoid polyps.
      Our feasibility study suggests that CNN assistance during live colonoscopy will result in fewer missed polyps. However, extrapolation to real-time use is limited by several factors, including unknown effects of the CNN on inspection behavior by the colonoscopist. Another limitation derives from the anonymized and de-identified nature of the videos, which excluded information about the indications for colonoscopy or the histology of polyps. CNN performance can vary by indication (screening vs surveillance).
      Polyp histology is especially relevant for added time and pathology costs. Time spent for polypectomy has “added value” whenever a true-positive result is precancerous, malignant, or relevant for calculating surveillance interval cancer and/or ADR. However, if use of the CNN results in polypectomies of clinically irrelevant lesions, the added time and unnecessary pathology costs would be unacceptable. Future randomized studies could directly address the overall value (quality vs cost) of the CNN by examining its effects on colonoscopy time, pathology costs, ADR, polyps per procedure, surveillance-relevant polyps per procedure, and surveillance-irrelevant polyps per procedure (normal, lymphoid aggregates, etc).
      Live use of the CNN can lengthen colonoscopy procedure times owing to second looks at false-positive findings and additional polypectomies. Time to assess a false-positive finding will likely average less than 5 seconds at an estimated FPR of fewer than 8 per colonoscopy. This relatively minor time cost could be lowered with further optimization of detection accuracy (eg, more training data), user interface (eg, color selection, sound effects), and simultaneous running of accurate optical pathology artifical intelligence algorithms.
      Although our results were obtained using Olympus endoscopes, which have a 70% endoscope market share, we expect that the proposed method will work with endoscopes from other vendors with little additional tuning of the algorithm. This is consistent with hundreds of experiments reported in the literature on “transfer learning” and our extensive experience with other biomedical imaging problems.
      • Wang J.
      • Fang Z.
      • Lang N.
      • et al.
      A multi-resolution approach for spinal metastasis detection using deep siamese neural networks.
      • Wang J.
      • Ding H.
      • Azamian F.
      • et al.
      Detecting cardiovascular disease from mammograms with deep learning.
      • Chang P.
      • Su L.
      • Baldi P.
      • et al.
      Deep learning convolutional neural networks accurately classify genetic mutations in gliomas.
      Our proposed method shows great promise in helping to close the gap between ADR and true adenoma prevalence, especially for the colonoscopist with a low ADR. By meeting the constraints of high accuracy and real-time performance using standard personal computers, this is the first reported polyp-detection artificial intelligence application ready for real-time validation studies.

      Conclusion

      We built a state-of-the-art polyp detection and localization model using deep learning that is easily capable of operating in real-time conditions (processing 1 frame in 10 ms). We detected the presence of polyps in a frame with an accuracy of 96.4% and an AUC of 0.991 using a CNN that was first trained on the ImageNet corpus of natural images and then retrained on our polyp database. A small adaptation of the model enabled it to localize polyps to within a bounding box with a state-of-the-art Dice/F1 score of 0.83 with a processing time of only 10 ms per frame. When overlaid on colonoscopy videos, the algorithm identified all polyps found by expert viewers (ADR > 50%) and additional polyps missed at expert review of non-overlaid videos. Therefore, we believe that, when running live during colonoscopy, this model will prompt more careful inspection and discovery of additional polyps. Thus, it is well positioned for validation in prospective trials to test effectiveness for improving ADR and lowering the adenoma miss rate. Furthermore, there is no reason to believe that the same methods, with the proper adjustments and training sets, could not work to tackle other real-time needs in endoscopy.

      Acknowledgments

      We thank Anish Patel, Junhee Kim, Gregory Albers, and Kenneth Chang for their support on this project. We also acknowledge a hardware gift from NVDIA and thank Yuzo Kanomata for computing support.

      Supplementary Material

       Architectures, Training, and Evaluation

       Neural Network Architectures

      We performed and evaluated experiments separately for polyp detection and polyp localization for clarity and to exclude confounding factors. In a use-case scenario, these can be performed by a single CNN model simultaneously. Polyp detection was implemented as a binary classification task of images by predicting the probability of whether an image would contain at least 1 polyp.
      For localizing polyps, we used an approach that leaves the model’s run time nearly unchanged to perform localization in real time. The neural network was trained to directly predict the center coordinates and the height and width of a bounding box that covers the polyp(s) in an image. An efficient approach to handling multiple polyp appearances in a single image is to predict not just 1 but several bounding boxes at once. This technique is a central aspect of “you only look once” (YOLO). In YOLO, the network has outputs spaced over a regular grid of the image (eg, 8 × 8), and each output predicts a confidence that an object of interest is within its grid cell and the object’s location and size (ie, 5 values are predicted). This allows the detection of multiple polyps simultaneously and incurs only a negligible computational overhead compared with a model that can predict the location of only a single object. We experimented with a variation of YOLO for better localizing single objects in an image, which we named the “internal ensemble.” For this all fully connected layers are replaced with convolutional layers. The last convolutional layer has 5 filters maps and will intrinsically have its outputs spaced over a grid over the input image (eg, 7 × 7). Each grid cell predicts its confidence (sigmoid unit), the position of the polyp relative to the grid cell center, and its size. The overall model output is the weighted sum of all 49 (7 × 7) adjusted position and size predictions, weighted with the predicted confidences. Hence, the model learns to use the confidence predictions (and thus weights) in an optimal way.
      Supplementary Table 1 presents the CNNs that were trained for polyp detection and localization. The convolutional layers of PI-CNN1–3 in Supplementary Table 1 are pre-initialized with weights that were obtained by training a closely related model on the ImageNet challenge dataset
      • Russakovsky O.
      • Deng J.
      • Su H.
      • et al.
      ImageNet large scale visual recognition challenge.
      —these model architectures are based on the VGG16,
      • Simonyan K.
      • Zisserman A.
      Very deep convolutional networks for large-scale image recognition.
      VGG19,
      • Simonyan K.
      • Zisserman A.
      Very deep convolutional networks for large-scale image recognition.
      and ResNet50
      • He K.
      • Zhang X.
      • Ren S.
      • et al.
      Deep residual learning for image recognition.
      models, respectively. We retrained all layers and weights of these CNNs on our polyp dataset. Networks NPI-CNN1 and -2 are not pre-initialized this way, but rather with random weights and thus trained from “scratch.” However, they are smaller to decrease overfitting and thus faster than the PI models. The weights of all layers of NPI models and those of all fully connected layers of PI models were initialized with an adaptively scaled normal distribution.
      • He K.
      • Zhang X.
      • Ren S.
      • et al.
      Delving deep into rectifiers: surpassing human-level performance on ImageNet classification.
      We experimented with a variety of architectures for non–pre-trained CNNs other than the ones presented in Supplementary Table 1. Although the reported networks had the lowest validation-set classification error rates, it should be noted that many alternative architectures reached very similar validation- and test-set accuracies. The number of layers, filters per layer, and maximum pooling and subsampling operations was chosen to keep the computational cost of evaluating the model low and to never pool by more than a factor of 2 between convolutional operations because this would negatively affect detection accuracy when using small filters.
      The following list provides overview of the different layer types commonly used in deep learning:
      • Convolutional layers convolve their input feature maps with a set of trainable filters to extract features. A sequence of convolutional layers extracts a hierarchy of features from implicitly increasingly large regions of the input image, with larger regions for layers higher in the hierarchy (further away from the input). The filter weights of a given layer are shared and used across the entire input field, which implies that convolutional layers are implicitly heavily regularized and usually contribute a relatively small amount of trainable weights compared with fully connected layers. Convolutions are at the core of most computer-vision and signal-processing models.
      • Fully connected layers correspond to a matrix–matrix multiplication with the output of the previous layer, where every entry in the matrix is a trainable parameter.
      • Batch normalization layers are introduced to prevent the distribution of each layer’s outputs from changing during training and have been shown to speed up learning and improve classification performance in some cases.
        • Ioffe S.
        • Szegedy C.
        Batch normalization: accelerating deep network training by reducing internal covariate shift.
        They normalize feature maps such that each resulting feature has 0 mean and unit standard deviation. The mean and standard deviation for the z-score normalization are estimated with a moving average during training.
      • Nonlinear activation functions are applied to the output of virtually every layer in the network. Without these, most deep network architectures would be mathematically equivalent to a simpler shallow linear model. Note that operations such as convolution, batch normalization, and matrix–matrix multiplication are intrinsically linear.
      • Maximum pooling layers summarize the outputs of neighboring regions in the same feature map. The output of a maximum pooling unit corresponds to the maximum of its input features taken over a certain region. In this study, the output was generated for every other location (ie, with stride 2) by considering the 2 × 2 neighborhood region of the location, denoted as p in Supplementary Table 1.

       Training

      When training a machine learning model to perform regression, one typically optimizes the L2 loss because it is the most natural loss function for general regression tasks for parameters such as polyp size and location. However, in this case, it implies treating the location coordinates and bounding-box dimensions independently from each other. This might not necessarily be optimal because highly accurate bounding-box width and height predictions are useless if the location predictions are bad. Thus, we also experimented with optimizing the Dice coefficient directly using a custom loss function, because it is a true measure of segmentation performance.
      All weights of polyp localization models were initialized with weights of the corresponding trained polyp-detection CNN to accelerate training and facilitate convergence. The neural networks were trained with the ADAM optimizer.

      Kingma DP, Ba JL, Adam: A Method for Stochastic Optimization. Proceedings of the 3rd International Conference on Learning Representations (ICLR) 2015.

      We used a batch size of 16 and optimized the initial learning rate for each architecture individually. We decreased the learning rate automatically when the validation accuracy did not improve for 10 consecutive epochs and stopped training after it had not improved for 50 epochs. In our experiments, we evaluated model performance by applying a 7-fold cross-validation procedure, because this decreases the variance of performance estimates compared with a simple single split. In each cross-validation run, the dataset was split into a different training, development, and test subset. The union of all test sections of the cross-validation splits recovers the original dataset (ie, the test sections are mutually exclusive). The CNN was trained on the training segment, whereas the development segment was used to monitor the CNN’s accuracy during training for early stopping and hyper-parameter optimization. All reported results were obtained on the test segments, which was left untouched during the training process. Furthermore, the weights of all CNNs were independently retrained from their initial configuration for each data split.

       Performance Evaluation and Metrics

      For evaluation of our (binary) classification results, we relied mainly on 2 metrics: accuracy and AUC. Machine learning models for classification typically predict a value that corresponds to a class probability. For a given threshold, performance is determined by the number of true positives (TP), true negatives (TN), false positives (FP), and false negatives (FN). Other quantities of interest such as accuracy, true positive rate (TPR; also called sensitivity or recall), and FPR can be easily obtained (see Baldi and Brunak
      • Baldi P.
      • Brunak S.
      Assessing the accuracy of prediction algorithms for classification: an overview.
      for more details):
      Accuracy=TP+TNTP+FP+FN+TNTPR=Sensitivity=TPP=TPTP+FNFPR=FPN=FPFP+TN


      In contrast, the AUC is invariant to the choice of threshold; it is the area under the receiver operating characteristic curve. The receiver operating characteristic is obtained by plotting the TPR against the FPR for all thresholds in the range [0, 1]. Thus, the best possible AUC is 1.
      To evaluate polyp localization and segmentation performance, we used the commonly used Sørensen–Dice coefficient D (also known as F1 score), which compares the relative area of overlap between the predicted bounding box (A) and the ground-truth (B) bounding box:
      D(A,B)=2|AB||A|+|B|


      The Dice coefficient is equivalent to Jaccard index J and also is known as the “intersection over union” (IoU) or Tanimoto distance. They can be trivially computed from each other using these relations:
      D=2J1+JJ=IoU=D2D


      Polyp Detection

      Supplementary Figure 2 shows the receiver operator characteristic curve for the 2 NPI models and the 3 PI CNNs when trained and evaluated on the 8,641 colonoscopy images (shown are test-set predictions for a single cross-validation split of the 5 models). The clinically most relevant section is the one with low FPRs (ie, high specificity), because excessive false alerts would be distracting and eventually be ignored by the colonoscopist. The 3 PI models are clearly better than the NPI models in this regime.

      Colonoscopy Video Study With Expert Colonoscopists

      The following tables present a detailed breakdown of results of expert review of videos with and without CNN assistance compared with CNN alone. The tables list polyp characteristics (size, shape, and location). Supplementary Table 2 also presents the confidence of the expert reviewer on the presence of the polyp and the number of polyps in 8 contiguous frames predicted by the CNN as “positive” (all >50% confidence to contain a polyp).

       Video Study 1

       Video Study 2

      Supplementary Table 4 presents an expert review of 11 deliberately challenging videos (withdrawal of colonoscope without closing in on already seen polyps).

       Statistics of Number of Polyps Per Video

       Effect of Post-Processing on Sensitivity and Specificity in Expert Video Study

      The CNN predictions on videos were filtered by a window size of 8 consecutive frames to remove likely false-positive findings. Supplementary Figure 1 shows the effect of different settings of this window size on the sensitivity and specificity of CNN predictions at a fixed threshold (on data from the first expert video study). A CNN prediction for any given frame is counted as positive if it is part of at least 1 window of consecutive predictions greater than 40% predicted confidence for polyp presence; otherwise, it is counted as negative. We found that a window size of 8 results in a well-balanced ratio of sensitivity to specificity.
      Figure thumbnail fx2
      Supplementary Figure 1Effect of the size of the filtering window on the sensitivity and specificity of CNN predictions using a fixed threshold of 0.4.
      Figure thumbnail fx3
      Supplementary Figure 2Receiver operator characteristic curve for all 5 CNN architectures trained on subsets of the 8641 colonoscopy images. Results obtained on the test splits of the 8641 colonoscopy images.
      Supplementary Table 1Summary of Architectures of Deep CNNs Used in This Study
      ModelInitial weightsArchitecture
      NPI-CNN164c-p-64c-p-128c-p-128c-p-256c-p-245c-p-512fc-512fc
      NPI-CNN224c-bn-p-24c-bn-p-24c-bn-p-Incept(16)-bn-Incept(16)-64c-p-64c-768fc-768fc
      PI-CNN1VGG1664c-64c-p-128c-128c-p-(256c)3-p-(512c)3-p-(512c)3-p-1560fc-1560fc
      PI-CNN2VGG1964c-64c-p-128c-128c-p-(256c)4-p-(512c)4-p-(512c)4-p-1560fc-1560fc
      PI-CNN3ResNet5064c-bn-p-Cb-(Ib)2-Cb-(Ib)3-Cb-(Ib)5-Cb-(Ib)2-p-1024fc
      NOTE. The first layer is the leftmost entry and the last hidden layer is the rightmost entry per row. Exponentiation means repetition (eg, [256c]3 corresponds to 3 convolutional layers with 256 filters each). Ib and Cb are structures containing 6 and 8 layers, respectively (for details, see Grave et al
      • Graves A.
      • Mohamed A.R.
      • Hinton G.
      Speech recognition with deep recurrent neural networks.
      ). Convolutional filter sizes are typically 3 × 3.
      bn, batch-normalization layer; c, convolutional layer; fc, fully connected layer; Incept, inception module
      • Wang J.
      • Fang Z.
      • Lang N.
      • et al.
      A multi-resolution approach for spinal metastasis detection using deep siamese neural networks.
      ; p, maximum pooling layer.
      Supplementary Table 2List of All Unique Polyps Found in the First Set of 9 Videos, With Their Location and Estimated Size
      Expert polyp encounterPolyp IDCNN alone (8 contiguous frames)Polyp characteristics
      Without CNNWith CNNRemoved or missedCNN+CNNCNN+, %Size (mm)ShapeLocationConfidence
      NoNoNANA270828,1469NANANANA
      NoYesMissed
      Missed polyps are those that were found with CNN assistance but not by expert review of videos without the CNN overlay.
      VID-1-3601002SessileHepatic flexureLow
      VID-1-5501004SessileTransverseHigh
      VID-3-3501002SessileRectumHigh
      VID-4-41701007SessileAscendingHigh
      VID-5-2601002SessileDescendingLow
      VID-6-3301002SessileDescendingLow
      VID-7-21201006SessileDescendingLow
      VID-7-6801006SessileDescendingLow
      VID-8-2101002SessileCecumLow
      Mean6.80.01003.7
      YesYesNot removedVID-1-2211972SessileHepatic flexureHigh
      VID-1-471946SessileAscendingHigh
      VID-2-1655932SessileSigmoidHigh
      VID-3-4601001SessileRectumLow
      VID-4-212801003SessileAscendingHigh
      VID-6-226261SessileAscendingHigh
      VID-7-12401003SessileAscendingHigh
      VID-7-516401003SessileSigmoidHigh
      Mean52.11.588.62.6
      YesYesRemovedVID-1-114978666SessileHepatic flexureHigh
      VID-1-690238010FlatAscendingHigh
      VID-1-7866943SessileDescendingHigh
      VID-1-820035852SessileRectumHigh
      VID-3-1244862SessileSigmoidHigh
      VID-3-21517952SessileSigmoidHigh
      VID-3-520101002SessileSigmoidHigh
      VID-4-1624159815FlatAscendingHigh
      VID-4-31024961SessileAscendingHigh
      VID-5-12564986SessileDescendingHigh
      VID-6-12,1882808910SessileAscendingHigh
      VID-7-323829895SessileDescendingHigh
      VID-7-429017945SessileSigmoidHigh
      VID-7-767563925SessileSigmoidHigh
      VID-8-18510891SessileCecumHigh
      VID-8-313917892SessileDescendingHigh
      VID-8-41266953SessileDescendingHigh
      VID-9-122176746SessileDescendingHigh
      VID-9-10868924SessileDescendingHigh
      VID-9-1110039720SessileSigmoidHigh
      VID-9-21967972SessileDescendingHigh
      VID-9-35301004SessileDescendingHigh
      VID-9-42401004SessileDescendingHigh
      VID-9-56701004SessileCecumHigh
      VID-9-6513954SessileCecumHigh
      VID-9-73301002SessileCecumHigh
      VID-9-81243983SessileCecumHigh
      VID-9-99201004SessileDescendingHigh
      Mean238.224.992.44.9
      a Missed polyps are those that were found with CNN assistance but not by expert review of videos without the CNN overlay.
      Supplementary Table 3Contingency Analysis Assessing Performance of CNN and CNN-Assisted Expert Review of the 9 Colonoscopy Videos
      Expert + CNNCNN alone
      PolypNon-polyp
      Polyp48,9113,577Sensitivity0.93
      Non-polyp17,407236,728Specificity0.93
      PPV0.74
      NPV0.985
      NOTE. Analysis is based on comparing expert annotations (on CNN-overlaid videos) and CNN predictions on a “frame-by-frame” basis across videos and assumes CNN-assisted expert review is the “gold standard.”
      NPV, negative predictive value; PPV, positive predictive value.
      Supplementary Table 4List of All Unique Polyps Found in Second Set of 11 Videos and Their Location and Estimated Size
      Polyp IDSize (mm)ShapeRemoved?Location
      10-13SessileYesCecum
      10-22SessileYesCecum
      10-34FlatYesCecum
      10-43SessileYesSigmoid
      10-55SessileYesRectum
      10-64FlatYesRectum
      11-12SessileYesCecum
      11-22SessileYesCecum
      11-32SessileYesCecum
      11-45SessileYesAscending
      11-53SessileYesAscending
      11-62SessileYesAscending
      11-72SessileYesAscending
      11-83SessileYesAscending
      11-91SessileNoAscending
      11-101SessileNoAscending
      11-113SessileYesAscending
      11-123SessileYesAscending
      11-133SessileYesAscending
      11-144SessileYesAscending
      11-152SessileNoAscending
      11-162SessileNoAscending
      11-173SessileYesAscending
      11-182SessileYesAscending
      11-192SessileYesAscending
      11-203SessileYesAscending
      11-212SessileYesAscending
      11-223SessileYesAscending
      11-233SessileYesHepatic flexure
      11-242SessileYesHepatic flexure
      11-253SessileYesHepatic flexure
      11-264SessileYesHepatic flexure
      11-271SessileNoHepatic flexure
      11-282SessileNoTransverse
      11-292SessileNoTransverse
      11-301SessileNoSplenic flexure
      11-312SessileNoSplenic flexure
      11-327PedunculatedYesDescending
      12-12SessileYesCecum
      12-21SessileYesAscending
      15-12SessileNoAscending
      16-11SessileNoCecum
      16-22SessileYesHepatic flexure
      16-34FlatYesSigmoid
      16-41SessileNoSigmoid
      17-12SessileNoDescending
      17-24SessileYesDescending
      18-12SessileNoAscending
      18-27SessileYesHepatic flexure
      18-31SessileYesSplenic flexure
      19-13SessileNoDescending
      20-12SessileYesCecum
      20-21SessileYesHepatic flexure
      20-32SessileYesTransverse
      20-44SessileYesTransverse
      20-56SessileYesTransverse
      20-64SessileYesSplenic flexure
      20-72SessileYesSplenic flexure
      20-83SessileYesSplenic flexure
      20-93SessileYesDescending
      20-103SessileYesDescending
      20-112SessileYesDescending
      20-125SessileYesDescending
      20-133SessileYesDescending
      20-143SessileYesDescending
      20-154SessileYesDescending
      20-163SessileYesDescending
      20-172SessileYesDescending
      20-184SessileYesSigmoid
      20-193SessileYesSigmoid
      20-204SessileYesSigmoid
      20-214SessileYesSigmoid
      20-222SessileYesSigmoid
      Supplementary Table 5Total Number of Polyps Found per Video in 20 Videos
      Video IDPolyps, n
      18
      21
      35
      44
      52
      63
      77
      84
      911
      106
      1132
      122
      130
      140
      151
      164
      172
      183
      191
      2022
      Author names in bold designate shared co-first authorship.

      References

      1. American Cancer Society. Cancer facts and figures 2016. Available at: https://www.cancer.org/research/cancer-facts-statistics/all-cancer-facts-figures/cancer-facts-figures-2016.html. Accessed September 3, 2018.

        • Strum W.B.
        Colorectal adenomas.
        N Engl J Med. 2016; 374: 1065-1075
        • Russell L.B.
        • Kuntz K.M.
        • Lansdorp-Vogelaar
        • et al.
        A systematic comparison of microsimulation models of colorectal cancer: the role of assumptions about adenoma progression.
        Med Decis Mak. 2011; 31: 530-539
        • Winawer S.J.
        • Zauber A.G.
        • Ho M.N.
        • et al.
        Prevention of colorectal cancer by colonoscopic polypectomy.
        N Engl J Med. 1993; 329: 1977-1981
        • Patel S.G.
        • Ahnen D.J.
        Prevention of interval colorectal cancers: what every clinician needs to know.
        Clin Gastroenterol Hepatol. 2014; 12: 7-15
        • Pohl H.
        • Robertson D.J.
        Colorectal cancers detected after colonoscopy frequently result from missed lesions.
        Clin Gastroenterol Hepatol. 2010; 8: 858-864
        • Leufkens A.M.
        • van Oijen M.G.H.
        • Vleggaar F.P.
        • et al.
        Factors influencing the miss rate of polyps in a back-to-back colonoscopy study.
        Endoscopy. 2012; 44: 470-475
        • Corley D.A.
        • Jensen C.D.
        • Marks A.R.
        • et al.
        Adenoma detection rate and risk of colorectal cancer and death.
        N Engl J Med. 2014; 370: 1298-1306
        • Than M.
        • Witherspoon J.
        • Shami J.
        • et al.
        Diagnostic miss rate for colorectal cancer: an audit.
        Ann Gastroenterol. 2014; 28: 94
        • Anderson J.C.
        • Butterly L.F.
        Colonoscopy: quality indicators.
        Clin Transl Gastroenterol. 2015; 6: e77
        • Kaminski M.F.
        • Wieszczy P.
        • Rupinski M.
        • et al.
        increased rate of adenoma detection associates with reduced risk of colorectal cancer and death.
        Gastroenterology. 2017; 153: 98-105
      2. GI Quality Measures for 2017 Released in MACRA Final Rule. Available at: http://partner.gastro.org/gi-quality-measures-for-2017-released-in-macra-final-rule. Accessed September 3, 2018.

        • Bond A.
        • Sarkar S.
        New technologies and techniques to improve adenoma detection in colonoscopy.
        World J Gastrointest Endosc. 2015; 7: 969
        • Hassan C.
        • Senore C.
        • Radaelli F.
        • et al.
        Full-spectrum (FUSE) versus standard forward-viewing colonoscopy in an organised colorectal cancer screening programme.
        Gut. 2017; 66: 1949-1955
        • Waldmann E.
        • Britto-Arias M.
        • Gessl I.
        • et al.
        Endoscopists with low adenoma detection rates benefit from high-definition endoscopy.
        Surg Endosc. 2015; 29: 466-473
        • Schmidhuber J.
        Deep learning in neural networks: an overview.
        Neural Netw. 2015; 61: 85-117
        • Baldi P.
        • Chauvin Y.
        Neural networks for fingerprint recognition.
        Neural Comput. 1993; 3: 402-418
        • Simonyan K.
        • Zisserman A.
        Very deep convolutional networks for large-scale image recognition.
        arXiv. 2014; (1409.1556)
        • He K.
        • Zhang X.
        • Ren S.
        • et al.
        Deep residual learning for image recognition.
        IEEE Conf Comput Vis Pattern Recogn. 2016; : 770-778
        • Redmon J.
        • Divvala S.
        • Girshick R.
        • et al.
        You only look once: unified, real-time object detection.
        Proc IEEE Conf Comput Vis Pattern Recogn. 2016; : 779-788
        • Szegedy C.
        • Liu W.
        • Jia Y.
        • et al.
        Going deeper with convolutions.
        Proc IEEE Conf Comput Vis Pattern Recogn. 2015; : 1-9
        • Graves A.
        • Mohamed A.R.
        • Hinton G.
        Speech recognition with deep recurrent neural networks.
        Acoustics Speech Signal Process. 2013; : 6645-6649
        • Wu Y.
        • Schuster M.
        • Chen Z.
        • et al.
        Google’s neural machine translation system: bridging the gap between human and machine translation.
        arXiv. 2016; (1609.08144)
        • Silver D.
        • Huang A.
        • Maddison C.
        • et al.
        Mastering the game of Go with deep neural networks and tree search.
        Nature. 2016; 529: 484-489
        • Wu L.
        • Baldi P.
        Learning to play GO using recursive neural networks.
        Neural Netw. 2008; 21: 1392-1400
        • Baldi P.
        • Sadowski P.
        • Whiteson D.
        Searching for exotic particles in high-energy physics with deep learning.
        Nat Commun. 2014; 5: 4308
        • Shimmin C.
        • Sadowski P.
        • Baldi P.
        • et al.
        Decorrelated jet substructure tagging using adversarial neural network.
        Physical Rev D. 2017; 96: 074034
        • Fooshee D.
        • Mood A.
        • Gutman E.
        • et al.
        deep learning for chemical reaction prediction.
        Mol Syst Des Eng. 2017; (10.1039/C7ME00107J)
        • Di Lena P.
        • Nagata K.
        • Baldi P.
        Deep architectures for protein contact map prediction.
        Bioinformatics. 2012; 28: 2449-2457
        • Wang J.
        • Fang Z.
        • Lang N.
        • et al.
        A multi-resolution approach for spinal metastasis detection using deep siamese neural networks.
        Comput Biol Med. 2017; 84: 137-146
        • Wang J.
        • Ding H.
        • Azamian F.
        • et al.
        Detecting cardiovascular disease from mammograms with deep learning.
        IEEE Trans Biomed Imaging. 2017; 36: 1172-1181
        • Chang P.
        • Su L.
        • Baldi P.
        • et al.
        Deep learning convolutional neural networks accurately classify genetic mutations in gliomas.
        AJNR Am J Neuroradiol. 2018; 39: 1201-1207
        • Esteva A.
        • Kuprel B.
        • Novoa R.
        • et al.
        Dermatologist-level classification of skin cancer with deep neural networks.
        Nature. 2017; 542: 115-118
        • Baldi P.
        Deep learning in biomedical data science.
        Annu Rev Biomed Data Sci. 2018; 1: 181-205
        • Russakovsky O.
        • Deng J.
        • Su H.
        • et al.
        ImageNet large scale visual recognition challenge.
        Int J Comput Vis. 2015; 115: 211-252
        • Ioffe S.
        • Szegedy C.
        Batch normalization: accelerating deep network training by reducing internal covariate shift.
        arXiv. 2015; (1502.03167)
        • Szegedy C.
        • Vanhoucke V.
        • Ioffe S.
        • et al.
        Rethinking the inception architecture for computer vision.
        Proc IEEE Conf Comput Vis Pattern Recogn. 2016; : 2818-2826
        • Chollet F.
        Keras.
        GitHub, San Francisco, CA2015
      3. Abadi M, Agarwal A, Barham P, et al. TensorFlow: Large-scale machine learning on heterogeneous systems. Available at: https://www.tensorflow.org/. Published 2015.

        • Hinton G.E.
        • Srivastava N.
        • Krizhevsky A.
        • et al.
        Improving neural networks by preventing co-adaptation of feature detectors.
        arXiv. 2012; (1207.0580)
        • Baldi P.
        • Sadowski P.
        The dropout learning algorithm.
        Artif Intell. 2014; 210: 78-122
        • Chatfield K.
        • Simonyan K.
        • Vedaldi A.
        • et al.
        Return of the devil in the details: delving deep into convolutional nets.
        arXiv. 2014; (1405.3531)
        • Bernal J.
        • Sánchez J.
        • Vilarino F.
        Towards automatic polyp detection with a polyp appearance model.
        Pattern Recogn. 2012; 45: 3166-3182
        • Ribeiro E.
        • Uhl A.
        • Häfner M.
        Colonic polyp classification with convolutional neural networks.
        IEEE 29th Int Symp Comput Based Med Syst. 2016; : 253-258
        • Soetikno R.
        • Friedland S.
        • Kaltenbach T.
        • et al.
        Nonpolypoid (flat and depressed) colorectal neoplasms.
        Gastroenterology. 2006; 130: 566-576
        • Park S.Y.
        • Sargent D.
        Colonoscopic polyp detection using convolutional neural networks.
        SPIE Med Imaging. 2016; (978528–978528)
        • Bernal J.
        • Tajkbaksh N.
        • Sánchez F.
        • et al.
        Comparative validation of polyp detection methods in video colonoscopy: results from the MICCAI 2015 Endoscopic Vision Challenge.
        IEEE Trans Med Imaging. 2017; 36: 1231-1249
      4. Olympus Annual Report 3. Available at: https://www.olympus-global.com/ir/data/annualreport/pdf/ar2017e_A3.pdf. Published 2018.